Description
What’s Included
- Sonomed E-Z Scan AB5500+ main unit
- A & B probe
- Power cable
- Foot pedal
- 90-day Digital Eye Center warranty
Condition & Warranty
Used. Tested and confirmed in excellent working order. Unit has been cleaned. Some signs of cosmetic use visible in product photos; mechanical and electronic function verified. Covered by a 90-day Digital Eye Center warranty on parts and workmanship.
Used Sonomed E-Z Scan AB5500+ Scan Ophthalmic Ultrasound Scanner
The Sonomed E-Z Scan AB5500+ is a portable, combined A-scan and B-scan ophthalmic ultrasound system built on over 30 years of Sonomed Escalon leadership in ophthalmic biometry. The 6.5″ color touch screen provides full alphanumeric data entry and image display in a compact 5.25 lb unit with adjustable legs that tilt from 0° to 60° for tabletop or slit-lamp-mounted workflows.
The A-scan delivers axial length measurements at electronic accuracy of ±0.023mm, with zone velocity gating for ACD, lens thickness, and vitreous chamber. Six built-in IOL power formulas — Holladay, Regression-II, Theoretic, Binkhorst, Hoffer-Q, and Haigis — cover the full range of surgical planning needs including cataract, dense cataract, aphakic, and pseudophakic eyes. Automatic calibration is performed via the built-in cylinder with no separate tools required.
The B-scan delivers 30 frames per second at 0.15mm tissue resolution with 100 dB logarithmic/linear/S-curve gain control and continuous zoom from 0.5× to 2.0× with joystick-controlled pan. RS-170 BNC output supports connection to a video printer or remote display. The unit is compatible with both contact and immersion probe techniques and accepts the included A & B probe for both modalities from a single instrument.
Technical Specifications
General
| Spec | Detail |
|---|---|
| Display | TFT Active Matrix Color LCD, 6.5″ diagonal, 640×480 px |
| Interface | Touch screen, full alphanumeric |
| Video output | RS-170 BNC (video printer / remote viewing) |
| Dimensions | 12.5″W × 3.25″H × 10.0″D |
| Weight | 5.25 lbs |
| Voltage | 100/120/220/240V, 50/60 Hz auto-sensed |
| Legs | Adjustable tilt 0°–60° |
A-Scan
| Spec | Detail |
|---|---|
| Probe | 10 MHz, focused, internal fixation light |
| Measurements | ACD, lens thickness, vitreous, axial length |
| IOL Formulas | Holladay, Regression-II, Theoretic/T, Binkhorst, Hoffer-Q, Haigis |
| Modes | Automatic & Manual; Cataract, Dense Cataract, Aphakic, Pseudophakic |
| Accuracy | Electronic ±0.023mm / Clinical ±0.1mm |
| Range | Auto: 18–33mm / Manual: 0.5–35mm |
| Calibration | Automatic (built-in calibration cylinder) |
B-Scan
| Spec | Detail |
|---|---|
| Probe | 10 MHz focused transducer, 30 frames/sec |
| Amplifier | 100 dB gain, Log/Linear/S-Curve + TVG controls |
| Magnification | Continuous zoom 0.5×–2.0× with joystick pan |
| Display resolution | 640×480 px color VGA |
| Tissue resolution | 0.15mm optimal |
| Measurements | Distance and area |